Confocal Microscopy

Confocal microscopy offers a number of significant improvements over conventional wide field fluorescence microscopy these benefits include; shallow depth of field, reduction of the out of focus glare and the ability to serially section the specimen.

When fluorescent specimens are imaged using widefield microscopy the whole specimen is bathed in light, all parts of the specimen are thus excited at the same time. Light originating from above and below the focal plane (secondary fluorescence) is captured along with light originating from it. In contrast, co focal microscopy uses a point light source (excitation pinhole), which is focused by the objective onto the specimen. Out of focus light is filtered out through the use of a second pinhole positioned in a conjugate plane in front of the detector. Thus, only light originating from the focal plane is detected

Fig 1: Light path for point scanning confocal Point scanning confocal microscopy

In this form of confocal microscopy, excitation light (laser) is focused onto a single point of the specimen and light returning from that point is collected using a photo multiplier tube (PMT).

The point of illumination is scanned across the specimen in a raster fashion (rectangular pattern of scanning line by line) to build up the final image. The typical arrangement of a point scanning confocal is described in figure 1.

Recent advances in point scanning confocal microscopy technology have improved the speed of image capture of these systems. In addition to the conventional galvanometer-driven scanning mirrors used in the scan heads of typical point scanning confocals, hybrid / tandem scanning confocals employ a second set of galvo scanning mirrors.

These second mirrors ‘resonate’ at a set frequency and can allow the rapid excitation and capture of images; full-frame image capture rates of ~25fps are achievable however; this rate can be increased by reducing the size of the image captured. The Nikon A1R point scanning confocal employs this technology.

Spinning disc confocal microscopy

Fig 2: Showing light path using spinning disc confocal microscopyThis form of confocal microscopy offers a number of advantages over the point scanning confocal microscope arrangement as outlined above – it also has a number of disadvantages. This technique allows for the rapid excitation and through a suitably fast CCD camera, acquisition of images. The optical arrangement of a typical spinning disk head is outlined in figure 2.

The spinning disk scanner uses a pair of aligned rotating disks; the top disk contains a spiral array of microlenses and the bottom one a similarly arranged array of 20,000 pinholes. Excitation light (in many cases, laser light) is expanded to illuminate the upper disk. Light-efficiency of the arrangement is increased through the presence of the microclines array, which focuses the light onto the pinholes of the lower disk. As the disk arrangement spins it traces concentric arcs of excitation light across the specimen; fluorescent light returns along the same path but is reflected by the diachronic mirror towards the CCD detector. Because the disks spin at high speed (>1800 rpm*), many fields of view (>300 frames per second*) can be excited within one camera exposure. The arrangement therefore collects light from multiple points simultaneously and in doing so, image capture rates can be fast (>50 frames per second) -although camera technology and signal strength may limit this. In addition to fast image capture, the spinning disk approach is ideal for the observation of photo-sensitive specimens as the sample is irradiated quickly and with less intense laser light than a conventional point-scanning confocal.

*Figures quoted for the CSU10 spinning disk unit as used on the Nikon / TIRF system.