The facility contains instrumentation as below;
This sorter has 4 lasers , they are Blue 488nm, Red 635nm,Violet 405nm and UV 355nm
The preferred fluorochromes for each detector are below, however many other combinations are possible , for advice please contact us .
(Laser-Band pass/Tolerance)
| 488-780/60 | 488-710/50 | 488-660/20 | 488-610/20 | 488-575/26 | 488-520/20 |
| PE-Cy7 | PerCP-eFluor 710 | PE-CF594 | Pe | FITC |
| 635-780/60 | 635-730/45 | 635-670/14 |
| APC-H7 | Alexa 700 | APC |
| 405-655/8 | 405-525/50 | 405-450/50 |
| BV650 | BV510 | BV421 |
| 355-675lp | 355-450/50 |
| Alexa 350 |
This instrument has 5 lasers, they are Blue 488nm, Red 635nm,Violet405nm, Green 532nm and UV 355nm
(Laser-Band pass/Tolerance)
| 488-780/60 | 488-710/50 | 488-660/20 | 488-610/20 | 488-575/26 | 488-520/20 |
| PE-Cy7 | PerCP-eFluor 710 | PE-CF594 | Pe | FITC |
| 635-780/60 | 635-730/45 | 635-670/14 |
| APC-H7 | Alexa 700 | APC |
| 405-655/8 | 405-525/50 | 405-450/50 |
| BV650 | BV510 | BV421 |
| 355-675lp | 355-450/50 |
| Alexa 350 |
| 532-710/50 | 532-585/42 |
| PeCy5.5 | Alexa 532 |
Two are 3 laser (Blue, Red, Violet) producing 8 colour capability
Two are 2 laser (Blue and Red) producing 6 colour capability
(Laser-Band pass/Tolerance)
| 488-780/60 | 488-670lp | 488-585/42 | 488-530/30 |
| PeCy7 | PerCP-eFluor 710 | PE | FITC |
| 635-780/60 | 635-660/20 |
| APC-H7 | APC |
| 405-510/50 | 450/50 |
| BV510 | BV421 |
This instrument has 3 Chanells 2 of which are capable of fluorescence detection
1-Bright filed, 2Green Fluorescence 3,Red Fluorescence
| Fluorescence parameter | Green Fluorescence | Red Fluorescence |
| Ex-Em/tolerance | 458- 525/50 | 530-585/lp |
| Viability kit Dead Cell Green | Viability kit Dead Cell Red |
A list of dyes that can be used with this instrument can be found here
The Bright field is used for cell counting
FACSFusion Sorter
To be updated
Fortessa X20
To be updated
Accessory instrumentation includes;
A comprehensive suite of flow Cytometry analysis software is also available on each of the two sites (Centre for life and Wm Leech Building ) to enable you to analyse acquired data quickly and efficiently, remote from the instruments.
The facility has a particular emphasis on flow cytometry education and we deliver both basic and advanced flow cytometry training. In addition.
All core facility staff strive to offer help and assistance to all users whether it's to assist in experimental set up, fluorochrome choice or any other issues that impact on the use of the facility.
We are happy to discuss any flow cytometry based work, or educational requirements from internal or external customers.
Software for the analysis of flow cytometry data can be found in both Laboratories situated on a stand alone analysis computer, you can not book these computers so please be aware that you may have to wait to use them
Software available includes